MaxQ 2000 Shaker
4-3
Thermo Scientific
Section 4
Operation
1. Press TIMER/ELAPSED membrane switch until TIMER and SET
TIME lights are illuminated. The HOURS or MINUTES light will
also light up at this point, depending on which option was previously
chosen.
2. Press HOURS/MINUTES membrane switch for desired timing mode.
3. Hold down appropriate arrow membrane switch in the TIME module
of the control panel, up or down, until desired timing cycle is set from
0.1 hour up to 999 hours, or from 0.1 minute to 999 minutes,
depending on which timing mode is chosen. SET TIME light will
illuminate.
4. Press START to begin timed shaking. Countdown will begin from
time set. TIMER and MINUTES or HOURS lights will illuminate
and timer will count down from time selected. An audible alarm will
sound at the end of the timing cycle and platform rotation will cease.
1. Press TIMER/ELAPSED membrane switch until ELAPSED light is
illuminated. The HOURS or MINUTES light will also light up at this
point depending on which option was previously chosen.
2. Press HOURS/MINUTES membrane switch for desired timing mode.
3. TIME display should show ‘0.0’. Press START to begin timed shaking.
Timer will begin to count up and will display accumulated time in
display window. Platform rotation will continue until the operator
presses the Stop button. The TIME display will flash when ELAPSED
time reaches 999 minutes or 999 hours.
The RS232 interface port is located on the left side rear of the shaker
cabinet and requires the use of a laptop or desktop computer running
Microsoft Windows 98 or newer operating system.
Note
RS232 Interface Port is for output only. Interface cables must not
exceed 9.8
ʼ
(3m) in length.
Setting Timer for
Timed Shaking
Setting Timer for
Continuous Shaking
RS232 Interface Port
RS232 Interface Port