47
Technical Service
G.2 Recommended Equipment Decontamination Methods
The following table lists the proper decontamination procedures for returning
Matrix Electronic pipettes to Thermo Fisher Scientific for service. Contact
customer service at 800-345-0206 (US) or +44(0) 161 486 2112 (Europe)
for a current equipment decontamination verification form.
Product/Component
Heat at
65 °C/
10 Hr.
70%/30%
IPA/Water
10%
Bleach
Ethylene
Oxide
Autoclave
121 °C/
15 min.
Multichannel Pipettes
Pipette Exterior
YES
Wipe
Wipe
YES
NO
Piston/Cylinder Assembly
YES
YES
NO
YES
YES
Tip Manifold
YES
YES
NO
YES
YES
Battery
NO
Wipe
NO
NO
NO
Power Supply
NO
Wipe
NO
NO
NO
Pipette Exterior
NO
Wipe
Wipe
NO
NO
Single Channel Electronic Pipettes
Pipette w/Battery Removed YES
Wipe
Wipe
YES
NO
Lower cylinder
YES
YES
YES
YES
YES
Piston (do not remove)
YES
YES
NO
YES
NO
Power Supply
NO
Wipe
NO
NO
NO
WARNING
: Do not expose batteries to temperatures above 60 °C.