29
Troubleshooting
E. TROUBLESHOOTING
Possible procedural and system problems, as well as recommended solutions.
If you cannot resolve the problem, call technical service at
800-345-0206
.
No power to the pipette.
Probable Cause
Recommended Solution
Battery is discharged.
Charge battery for at least one hour.
Double beep tone occurs.
Probable Cause
Recommended Solution
An illegal programming step
is being entered, such as
programming two fills that
exceed the maximum fill
volume.
Recheck the programming sequence.
Ensure you are not attempting to fill
above the maximum fill volume of
the pipette, or attempting to dispense
more than the total fill volume.
“ERR_1” is displayed. (Signifies one of the following conditions)
Probable Cause
Recommended Solution
Liquid is aspirated into the
pipette.
Disassemble and clean the pipette.
(See Section F.1 and/or F.2, for pipette
cleaning and autoclaving instructions).
Motor is defective or the flex
circuit assembly is damaged.
Contact Thermo Fisher Scientific’s
technical service at 800-345-0206