Tuning and Calibrating Automatically in the ESI/MS Mode
Setting Up the MS Detector in the Xcalibur Data System for Tuning and Calibration
Thermo Electron Corporation
LTQ XL Getting Started
49
3. Open the Tune Method file that stores the factory default tune settings
for low-flow ESI operation, as follows:
a. Choose
File > Open
to display the Open dialog box.
b. Browse for the folder C:\Xcalibur\methods. Select the file
AutoTune.LTQTune.
c. Click
Open
to open the file. Tune Plus downloads the Tune Method
parameters to the MS detector.
4. Examine the pre-tune ESI source settings as follows:
a. From the Instrument Setup toolbar, click the
API Source
button to
open the ESI Source dialog box. Verify that the settings in your
dialog box are the same as those shown in
b. Click
OK
to return to the Tune Plus window.
Figure 19.
ESI Source dialog box, showing the settings to start a typical low
flow experiment
5. Set the scan parameters for tuning and calibration, as follows:
Note
The Xcalibur data system contains customized tune files for
different applications in the folder
C:\Xcalibur\methods
, including
one for low flow LC/ESI/MS operation.