Replacing the CPU PROM & HART PROM
Procedure
Thermo Fisher Scientific
LevelPRO Installation Guide
D-5
Figure D–6.
Figure D–7.
Once the old PROM is removed, align the new one on the socket so that
the beveled corner of the PROM (shown in Figure D–7) matches the
beveled corner of the socket. Push gently down on the PROM until it
snaps into place in the socket.
If becomes misaligned or is hard to push,
stop and start over.
Once it snaps into the socket, it is installed.
After installing the detector, you will need to erase memory, configure the
gauge, and calibrate as described in the main body of this manual.