Thermo Scientific KingFisher Plant DNA Kit Instruction Manual
General Information
Elution step
Carryover of ethanol to the Elution Buffer causes impurities
in the Elution Buffer and may affect some downstream
applications. To remove traces of ethanol, make sure that
there is a wash step (e.g., washing without releasing the
beads) or the drying step before the elution step is long
enough. There is a delicate balance between complete
removal of the ethanol and loss of DNA.
The volume of the Elution Buffer can be modified
depending on the user requirements concerning
the purified DNA concentration. The final DNA
concentration in the Elution Buffer may increase if the
purified DNA is eluted into a smaller volume of the
buffer, but this can slightly reduce the overall DNA yield.
The modifications of the elution step must be done in
BindIt Software 3.2 and according to the volume ranges
suitable for the KingFisher instrument. The table below
indicates the available elution volumes of the KingFisher
instruments.
Table 6-1.
Available elution volumes of Thermo Scientific KingFisher
instruments
KingFisher instrument
Elution volume
KingFisher
20–200 µl
KingFisher mL
50–1000 µl
KingFisher Duo with 12-pin magnet head
30–130 µl
KingFisher Duo with 6-pin magnet head
200–5000 µl
KingFisher Flex with 96 deep well head,
elution in a KingFisher Flex 96 KF plate
50–150 µl
KingFisher Flex with 96 deep well head,
elution in a Microtiter deep well 96 plate
50–1000 µl
KingFisher Flex with 96 well head,
elution in a KingFisher Flex 96 KF plate
20–250 µl
KingFisher Flex with 24 deep well head
200–5000 µl
Elution step
Содержание KingFisher Plant DNA Kit
Страница 1: ...Thermo Scientific KingFisher Plant DNA Kit Instruction Manual Rev 1 1 ...
Страница 2: ......
Страница 3: ...Thermo Scientific KingFisher Plant DNA Kit Instruction Manual Rev 1 1 Cat no N11996 ...
Страница 10: ...Thermo Fisher Scientific 10 Thermo Scientific KingFisher Plant DNA Kit Instruction Manual Kit Content ...
Страница 18: ...Thermo Fisher Scientific 18 Thermo Scientific KingFisher Plant DNA Kit Instruction Manual Safety Information ...
Страница 39: ......
Страница 40: ...Thermo Fisher Scientific Oy Ratastie 2 P O Box 100 FI 01621 Vantaa Finland www thermoscientific com N11996 ...