3
Advanced Maintenance
Replacing Components of the Manifold
252
ISQ 7000 Hardware Manual
Thermo Scientific
Source Retainer Magnets, Magnet Yoke, Source Insertion Guide, Source Magnets
and Front Manifold O-ring
To replace the source retainer magnets, magnet yoke, source insertion guide, source
magnets, or front manifold o-ring on the manifold door
1. Shut down your system by following the steps in
“Shutting Down the ISQ 7000 System”
.
2. Use a T20 Torxhead screwdriver to remove the four screws around the manifold door.
Note
For information about replacing the tray alignment pin or the o-ring on the back
manifold plate, see
.
Note
For information about ordering these components, see the
ISQ and TSQ GC-MS
Spare Parts Guide
.
Note
You do not have to remove the ion source cartridge or the column. However, you
must vent the instrument.
Figure 263.
Opening the Manifold Door
Screws