3
Advanced Maintenance
Replacing Components of the Manifold
Thermo Scientific
ISQ 7000 Hardware Manual
249
Back Manifold Plate
To replace the back manifold plate
1. Shut down your system by following the steps in
“Shutting Down the ISQ 7000 System”
.
2. Use a T20 Torxhead screwdriver to loosen the top cover panel screw from the interior
front panel. Then slide the top cover panel toward the back of the instrument and lift it
off.
3. Disconnect the power cables from the back of the ISQ instrument.
4. Use a T20 Torxhead screwdriver to remove the two screws holding the power supply
system to the top of the back panel.
5. Grasp the power supply system and lift it out of the instrument. It fits snug, so you may
have to shake it as you lift it out
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering this
component.
Note
You do not have to remove the ion source cartridge or column. However, you must
vent the instrument.
Figure 260.
Disconnecting the Power Cables
Power Cable
Connectors
Screws