Advanced Maintenance
Replacing Components of the Ion Source
210
ISQ 7000 Hardware Manual
Thermo Scientific
Lens Heater Block Grounding Strap
To replace the lens heater block grounding strap
1. Shut down your system by following the steps in
“Shutting Down the ISQ 7000 System”
.
2. Remove the analyzer tray from the instrument by following the instructions in
“Removing the Analyzer Tray from the ISQ 7000 Instrument”
3. Use a T10 Torxhead screwdriver to disconnect the screws holding the lens heater block
grounding strap to the lens heater block and analyzer tray.
4. Attach a new lens heater block grounding strap and reattach the screws.
5. Once the components are correctly attached to the analyzer tray, follow the instructions
in
6. Restart your system by following the instructions in
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering this
component.
Figure 198.
Replacing the Lens Heater Block Grounding Strap
Screws
Lens Heater
Block
Lens Heater Block
Grounding Strap