Advanced Maintenance
Replacing Components of the Ion Source
198
ISQ 7000 Hardware Manual
Thermo Scientific
Insulating Spacers
To replace the insulating spacers
1. Shut down your system by following the steps in
“Shutting Down the ISQ 7000 System”
.
2. Use a T20 Torxhead screwdriver to remove the four screws around the manifold door.
3. Twist off the source-repeller thumbscrews one at a time to prevent the repeller plate from
falling out.
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering this
component.
Figure 182.
Opening the Manifold Door
Figure 183.
Removing the Source-Repeller Thumbscrews
Screws
Source-Repeller
Thumbscrews