Alarm
Description
Recommended Action
Front/Rear FanX speed below
threshold YYYY.
Contact Technical Support.
Where X is the fan number and YYYY
is the low threshold speed for the
system fan.
FPGA failed to connect in emulate
mode.
Internal error.
Valve Link Failure.
OIA (On Instrument Analysis) is not
running.
FPGA temperature is above the
threshold of 70°C.
One or more of the system fans has
failed.
Boot drive exceeded 75%.
The boot drive is unexpectedly full.
RAID degraded one disk has failed.
One of the Torrent Server drives has
failed. System is still operational but
must be serviced soon.
RAID failure detected, more than one
disk failed.
More than one of the Torrent Server
drives has failed. System is no longer
operational.
/sw_results directory not mounted.
One or more of the drives used during
data collection has failed.
No GPU detected.
GPU has failed.
No GPU Driver detected.
Air Compressor failure.
Air compressor is either leaking or
has stopped working.
Nonfatal alarms
Instrument must be cleaned before
starting another run.
Maximum number of runs since last
clean has been exceeded.
Perform an instrument cleaning (see
“Manually clean the instrument“ on
page 14).
Ambient temperature is above the
threshold of 35°C.
Verify the temperature in the lab.
Contact Technical Support if the
temperature of the lab is within
normal operating temperature (20°C
to 30°C).
RAID degraded one disk has failed.
One of the Torrent Server drives has
failed. System is still operational but
must be serviced soon.
Contact Technical Support.
No Connectivity to Torrent Server.
Check that the Ethernet cable
connecting the instrument to the
server is properly connected. See
“Ion GeneStudio
™
S5 Food Protection
Instrument input and output
connections“ on page 11 for the
Ethernet port location.
Reconfigure the Torrent Server as
needed (see “Network Settings“ on
page 18.
No Connectivity to FTP server.
Reconfigure the FTP server as
needed, see “Network Settings“ on
page 18.
Appendix A
Touchscreen reference
Alarms, Notifications, and Events
A
36
Ion GeneStudio
™
S5 Food Protection Instrument User Guide