7.
Touch
Start run
to begin the run. When the run begins, the
Run
screen will
Note:
If the run needs to be stopped or paused before completion, touch
Stop
or
Pause
. If run parameters need to be edited during the run, pause the run,
enter changes, then touch
Resume
. When resumed, the run time will continue
from where it was paused.
8.
When the run is complete, the
Run Complete
screen will be displayed (see
Figure 13).
If the method used is a custom method that has not yet been saved, the
Run
Complete
screen will display a
Save as
option. If the method is a saved or
pre-programmed method, the
Save as
option will not be present. This screen
also displays an option to navigate to the
Home
screen or to repeat the method.
Constant current
Discontinuous buffer systems and, to a lesser extent, continuous systems increase
resistance during the run. With the current limiting setting on the PowerEase
™
Touch
350W Power Supply, the voltage will increase as resistance increases to satisfy
Ohm’s law (V=IR). If no voltage limit is set and a local fault condition occurs, such as
a poor connection, very high local resistance may cause the voltage to increase to
the maximum capacity of the power supply. This may lead to local overheating and
damage to the electrophoresis cell or create unsafe conditions.
Note:
When operating under constant current conditions, it is recommended to set a
voltage limit on the power supply slightly above the maximum expected for the run.
Constant Power
Power is a function of voltage and current P=IV. For a typical gel run at constant
power, the current will decrease during the run, causing the voltage to increase. The
power limiting function may be used when running sequencing gels to remove the
ammonium persulfate from the wells and to heat the gel to an optimal temperature for
DNA separations.
Custom methods
Select
Custom Methods
on the
Home
screen to bring up the
Custom Methods
screen. From here, the
Saved Methods
or
Method Type
screen can be selected (see
The
Saved Methods
screen displays methods that have been previously saved
(shown in alphabetical order). Selecting a method brings up the run screen for the
method, allowing the method to be run or edited. The saved method can be edited
by touching any of the run parameter fields.
The
Method Type
screen displays method types that can be entered (see Figure 5).
Chapter 2
Methods
Custom methods
2
PowerEase
™
Touch 350W Power Supply User Guide
13