Set up a run
1.
Switch the power switch located at the rear of the unit to the
On
position.
2.
Touch
Set up run
.
3.
In the
Method Type
screen touch either
Constant voltage
,
Constant current
,
or
Constant power
.
Figure 5 Method Type screens
4.
Edit the run parameters (Volts, Time, Amps, and Watts) by touching the
parameter on the screen to bring up the
Run parameter edit
screen.
Default maximum values are 300 V, 3,000 mA, and 350 W.
Note:
When using constant voltage or constant power, it is recommended
that the maximum default settings not be changed to avoid unwanted power
limitations during the run.
5.
Enter the parameter value required, then touch
Enter
to advance with the newly
changed run parameter displayed.
6.
(Optional)
Add or delete extra steps by touching
Actions
.
Chapter 2
Methods
Set up a run
2
12
PowerEase
™
Touch 350W Power Supply User Guide