System Description
Cooling System
Thermo Fisher Scientific
iCAP Q Operating Manual (P/N 1288090, Revision B)
3-19
Cooling System
Some components of the iCAP Q mass spectrometer, for example the
interface and the RF generated, must be water-cooled for operation. See
for the cooling water specification.
The cooling water must be pure and not contain additives. An in-line
filter is supplied with the instrument to guarantee consistent water
quality.
Connections for Water Supply
The iCAP Q system can be supplied with cooling water either from a
central system or from the optionally shipped recirculation water chiller
NESLAB ThermoFlex 2500. The connections for the cooling water are
shown in
Table 3-8.
Cooling water specification
Item
Specification
Supply rate
> 5.5 L/min
Temperature
15 to 25 °C, optimum at 21 °C
Pressure
0.25 to 0.55 MPa (2.5 to 5.5 bar)
Solid residual
< 50 μm particle size
pH
7 to 8
Fluid
distilled water
Labeled Components: 1=cooling water in, 2=cooling water out, 3=water filter
Figure 3-16.
Water cooling connection detail of left view of iCAP Q
1
2
3