P
Intended Use
iv
Heraeus Pico
17 / 21
Thermo Scientific
Intended Use
•
This centrifuge is a laboratory product designed to separate components by generation of a
relative centrifugal force. It separates human samples (e.g. blood, urine and other body fluids)
collected in appropriate containers, either alone or after addition of reagents or other additives.
•
As general-purpose centrifuge, it is designed to also run other containers filled with chemicals,
environmental samples and other non-human body samples.
•
Together with the hematocrit rotor 75003473 the centrifuge can be used as an in-vitro
diagnostics device to define the hematocrit value by centrifugation. Apply DIN 58933 to define
the hematocrit value.
•
The centrifuge is to be used for separating materials of different density or particle size
suspended in a liquid.
Maximum sample density at maximum speed: 1.2
Accident Prevention
g
cm
3
-----------
WARNING
If a hazardous situation occurs, turn off the power supply to the centrifuge
and leave the area immediately.
WARNING
In case of rotor failure the centrifuge can be damaged. The coolant can
escape. Ventilate the room well and leave it. Inform customer service.
WARNING
Plug the centrifuge only into sockets which have been properly grounded.
Prerequisite for the safe operation of the
Heraeus Pico 17 / 21
is a work environment in
compliance with standards, directives and trade association safety regulations and proper
instruction of the user.
The safety regulations contain the following basic recommendations:
•
Maintain a radius of at least 30 cm around the centrifuge.
•
Implementation of special measures which ensure that no one can approach the
centrifuge for longer than absolutely necessary while it is running.
•
The mains plug must be freely accessible at all times. Pull out the power supply plug or
disconnect the power supply in an emergency.