Sarasota 200 Ultrasonic Multipath
Flowmeter
Appendix 1 GAFA PC Software Page A1-13
Thermo Fisher Scientific
APPENDIX 1 GAFA PC SOFTWARE
•
The “Transducer Type” is identified by the Sarasota 200 and is for information only.
•
The normal transducer “Charge Mode” is “Single Shot” which consumes minimum power.
“Continuous” may be used for single path operation for convenience during setting-up.
•
The “HT Drive” is usually set to “Manual” for setting up in which case the “Drive Voltage”
and “Gain” are set manually. The effect of changing them is observed via the Waveform
Capture Screen. Once set up, “Auto Gain” may be selected.
•
“Shape Recognition” is used to reject distorted signals. It may be switched off for
diagnostic purposes.
•
Click on “Read Path Stats” to view the autogain statistics. This is for diagnostic purposes.
•
Click on “Write To Gauge” to confirm selections to the Sarasota 200 or “Cancel” to
abandon.