Four spare fuses are provided with this shaker and are taped to the
underside of the control panel plastic frame. The plastic frame is attached
to the cabinet by Velcro strips. The panel can be removed by grasping the
corners and pulling it off. There are also small indents located along the
edges of the panel, which accommodate a flat screwdriver blade.
The locations of the fuses are identified in Figure 1-16.
Figure 4-2.
Underside of Frame
1. Make sure the unit is turned off, unplugged, and the line cord removed
from the side of the cabinet. The platform should already be removed.
2. Remove the screws and lift off the cover plate. The mechanism will be
accessible as shown in Figure 4-3.
3. Clean the mechanism area with a 70% solution of alcohol. Everything
must be cleaned with special attention given to the area around the
motor.
4. Allow the unit to air-dry about one hour before reassembling.
5. The shaker is ready to be returned to service.
4-4
Orbital Shakers
Thermo Scientific
Section 4
Service
Spare Fuses
Cleaning Inside the
Cabinet (Major Spill)
Figure 4-3.
Behind Cover Plate