Enter the System Configuration by pressing and holding
the knob. After about 5 seconds a single you may remove
your finger from the knob. The display indicate "System
Config" on the first row, the "Temp Unit" will show on the
second row of the display. Turn the knob clockwise until
the "Preset Continue" show on the second row of the
display. If you do no action, the display will return to the
operation status.
Press the knob to enter the Preset Continue setting
interface.(Long press will return to operation menu.)
The "User OTP" show on the first row, "OFF" flash on
the second row. Turn the knob clockwise to change the
status. When "On" has reached, wait 10 seconds, the
display will return to operation menu indicating that it has
been set.
Temperature Probe
Calibration
This section applies only to the calibration of the
temperature probe system. The hot plate surface and the
motor speed cannot be calibrated. This method is only
applicable to units that have heating functionality and with
the probe connected.
Before calibrate the probe, connect the probe on the unit
and put the probe sensor into the temperature reference.
Enter the System Configuration by pressing and holding
the knob. After about 5 seconds a single you may remove
your finger from the knob. The display indicate "System
Config" on the first row, the "Temp Unit" will show on the
second row of the display. Turn the knob clockwise until
the "Calibrate Temp" show on the second row of the
display. If you do no action, the display will return to the
operation status. The option "Calibrate Temp" will only
show when there is a remote probe PT100 or PT1000
adopted.
Press the knob to enter the Calibrate Temp setting
interface. The "Calibrate Temp" show on the first row, the
measured temperature flash on the second row. If the data
is not matched the temperature you measure by external
thermo detector, you can turn the knob to calibrate. When
it’s matched, press the knob to complete the calibration.
When setting and/or viewing is complete, the knob can be
pressed to return to the previous operation menu.