Heating Specifications*
Top Plate Surface - Solid Ceramic
Models
7” x 7”
10” x 10”
Temperature range
30°C - 450°C (86°F - 842°F)**
30°C - 400°C (86°F - 752°F)**
Heat-up time to maximum temperature (unloaded top
plate)
5 minutes
7 minutes
Accuracy of the temperature display vs. the actual av-
erage temperature of a 2” diameter of setting area at
the center of the top plate (setpoint 100°C unloaded)
± 5.0°C
Temperature stability at the center of the top plate sur-
face (@ 100°C unloaded)
± 1.0°C
Typical accuracy of remote probe at user selected cal-
ibration temperature after calibration procedure
± 0.5°C
Temperature stability using remote probe (500 ml of
water in a 1000 ml flask at 70°C)
± 0.5°C
Top Plate Surface - Solid Aluminum
Models
7” x 7”
Temperature range
30°C - 300°C (86°F - 572°F)**
Heat-up time to maximum temperature (unloaded top
plate)
5 minutes
Accuracy of the temperature display vs. the actual av-
erage temperature of a 2” diameter of setting area at
the center of the top plate (setpoint 100°C unloaded)
± 5.0°C
Temperature stability at the center of the top plate sur-
face (@ 100°C unloaded)
± 1.0°C
Typical accuracy of remote probe at user selected cal-
ibration temperature after calibration procedure
± 0.5°C
Temperature stability using remote probe (500 ml of
water in a 1000 ml flask at 70°C)
± 0.5°C
Stirring Speed Specifications
Speed Range 50 to 1500 RPM(Maximum speed is dependent on the viscosity of the solution).
Stability of the stirring speed setpoint (600 ml of water in a 1000 ml glass flask)
± 2.0% at 1000 RPM
Top Plate Size
Max Recommended Flask Size
Max Weight on Top Plate
7” x 7”
4 liters
25 lbs
10” x 10”
6 liters
35 lbs
* The specification defined in this section is derived from testing data statistics of Thermo Fisher Scientific lab. We reserve
the right to make changes in design and specification without prior notice.
** As the top plate becomes dirty, the maximum temperature will decrease. To return the unit to its maximum temperature
performance, use a mild abrasive to remove stained areas.