Elite CTS Tester Operating Instructions

 
Manual Calibration

When the conductivity reading is outside calibration range of the automatic conductivity standards or when TDS or 
salinity is used, the tester will require manual adjustment.

1.  Repeat steps 1 to 4 from “Automatic Calibration for Conductivity”.
2.  Press 

 to manually adjust the value to the desired reading.

Note: 

The adjustment will decrease only, however the adjustment will eventually cycle to the  highest available value 

after decreasing by 40% of the initial value.
3.  Press 

 to accept and finish the calibration when the desired value is selected. 

4.  To abort calibration, press 

 to escape.

5. 

Once the calibration is finished and user has accepted the changes, measurement window will now show the 

calibrated reading.

Note: 

The auto conductivity standards are 84 µS/cm, 1413 µS/cm & 12.88 mS/cm.

 

CTS Measurement

1.  Press 

 to power the tester on if needed.

2.  Dip the electrode about 2 cm to 3 cm into the test solution. Stir and let the reading stabilize. The timer icon will 

blink during this time. Once the reading is stabilized, the timer stops blinking and   will appear to indicate the 
stability of the reading. 
 

CAUTION: 

Testing dry samples is not accurate and can lead to sensor damage or  

  

     breakage. Soils must be wet and free of particulates that may scratch the  

                    glass sensor. Excessive force into dry samples can cause glass breakage. 

3.  Note the CTS value or press 

 to freeze the reading. To release the reading, press 

 again.

4.  Press 

 to turn off tester. If key is not pressed for 8.5 minutes, the tester will automatically shut off to conserve 

batteries.

User Reset

Reset the CTS calibration to the user’s default settings by using the user reset function. Temperature user calibration 
is not affected by the user reset function.
1.  Press      to enter setup window. Scroll down by pressing 

 to select Reset. Press 

 to select Reset. The 

display shows User Reset and Fact Reset.

2.  Press 

 to select User Reset. 

3.  The display automatically shows No

 

and

 

Yes. Scroll down by pressing 

 to toggle between No

 

and

 

Yes

4.  Press 

 to confirm either No or Yes. The display shows 

the User Reset option with a  . 

 

COND

  

 µs/cm

      820 
    24.8 °C

Measure

Settings 

Temp Set

Key Info   

Scroll 

Accept 

EXIT 

Fact Reset

No

 

Yes

User Reset

Reset

 

CAL

  

µS/cm

        

    

844

    1413

 

CAL

   

   

Place the tester 
in the calibration 
STD to start 
or press   to 
escape.

 

CAL

    µS/cm

      

844

 820

 

COND

  

µS/cm

        

    

844 

     24.8 °C

Содержание Elite CTS

Страница 1: ...Elite CTS Tester Operating Instructions 68X001308 Rev 0 June 2017...

Страница 2: ...erating Instructions This instruction manual is applicable for Elite CTS Tester Replacing the Batteries The Elite CTS tester uses four AAA 1 5 V batteries 1 To remove the battery cover See Figure 1 Cl...

Страница 3: ...Press to enter setup window Scroll down by pressing to select Temp Set Press to select Temp set The display shows Set C F Temp Cal and Temp Coeff 2 Press to select Set C F Scroll down by pressing to t...

Страница 4: ...red off The conductivity value can be calibrated automatically or manually while the TDS salinity values require manual calibration The tester will begin in the measurement mode that was used when it...

Страница 5: ...imer stops blinking and will appear to indicate the stability of the reading CAUTION Testing dry samples is not accurate and can lead to sensor damage or breakage Soils must be wet and free of particu...

Страница 6: ...its protective cap For cup type electrodes remove the white plastic cup and insert to thoroughly clean viscous solutions Never scratch electrodes with a hard substance 2 For better performance soak th...

Страница 7: ...eed to change the electrode 1 With dry hands grip the ring with sensor facing you Twist the ring clockwise Save the ring for later use 2 Pull the old sensor module away from the tester 3 Align the fou...

Страница 8: ...to 1000 ppm 0 10 to 10 00 ppt TDS Factor 0 5 Resolution 0 1 ppm 1 ppm 0 01 ppt Relative accuracy 1 full scale Calibration points Up to 3 points TDS Factor 0 40 to 1 00 Selectable Salinity Salinity ra...

Страница 9: ...TSPIN Conductivity TDS salinity pocket tester with batteries ELITECTSCUP Conductivity TDS salinity pocket tester with batteries Conductivity Standard Solutions and Sachets ECCON100BT 100 S cm KCl 480...

Отзывы: