TDS Factor Setting
1. Press
to enter setup window. Scroll down by pressing
to select Settings.
2. Press
to select settings. The display shows TDS factor and Backlight.
3. Press
to select the default TDS factory setting or
to adjust the setting.
4. Press
to confirm the selection of the setting. The display shows the set TDS factor with a
.
Backlight Settings
1. Press
to enter setup window. Scroll down by pressing
to select Settings.
2. Press
to select Settings. The display shows TDS Factor and Backlight.
3. Scroll down by pressing to toggle between TDS Factor and Backlight. Press
to select Backlight.
4. The display shows ON and OFF. Scroll down by pressing to toggle between ON and OFF. Backlight ON
increases readability in low light conditions.
5. Press
to select the desired backlight option. The display shows the selected backlight option with a .
Temperature Settings
1. Press to enter setup window. Scroll down by pressing
to select Temp Set. Press
to select
Temp set. The display shows Set °C/°F, Temp Cal and Temp Coeff.
2. Press
to select Set °C/°F. Scroll down by pressing
to toggle between °C
and
°F
.
3. Press
to select tempareture unit. The display shows the selected temperature setting with a .
Temperature Calibration
1. Press to enter setup window. Scroll down by pressing
to select Temp Set.
2. Press
to select Temp Set. The display shows Set °C/°F, Temp Cal and Temp Coeff.
3. Scroll down by pressing to toggle between Set °C/°F, Temp Cal and Temp Coeff. Press
to select
Temp Cal.
4. The lower display shows the current measured temperature reading based on the last set offset and the upper
display shows the current measured temperature reading based on factory default calibration.
Measure
Settings
Temp Set
Reset
Key Info
Scroll
Accept
EXIT
TDS Factor
Backlight
ON
OFF
Backlight
OFF
Measure
Settings
Temp Set
Reset
Key Info
Scroll
Accept
EXIT
Backlight
Adjust
TDS Factor
0.71
0.66
TDS Factor
Measure
Settings
Reset
Key Info
Scroll
Accept
EXIT
Temp Cal
Temp Coeff
°C
°F
°C
Temp Set
Set °C/°F