Elite CTS Tester Operating Instructions
68X001308 Rev.0
June, 2017
Страница 1: ...Elite CTS Tester Operating Instructions 68X001308 Rev 0 June 2017...
Страница 2: ...erating Instructions This instruction manual is applicable for Elite CTS Tester Replacing the Batteries The Elite CTS tester uses four AAA 1 5 V batteries 1 To remove the battery cover See Figure 1 Cl...
Страница 3: ...Press to enter setup window Scroll down by pressing to select Temp Set Press to select Temp set The display shows Set C F Temp Cal and Temp Coeff 2 Press to select Set C F Scroll down by pressing to t...
Страница 4: ...red off The conductivity value can be calibrated automatically or manually while the TDS salinity values require manual calibration The tester will begin in the measurement mode that was used when it...
Страница 5: ...imer stops blinking and will appear to indicate the stability of the reading CAUTION Testing dry samples is not accurate and can lead to sensor damage or breakage Soils must be wet and free of particu...
Страница 6: ...its protective cap For cup type electrodes remove the white plastic cup and insert to thoroughly clean viscous solutions Never scratch electrodes with a hard substance 2 For better performance soak th...
Страница 7: ...eed to change the electrode 1 With dry hands grip the ring with sensor facing you Twist the ring clockwise Save the ring for later use 2 Pull the old sensor module away from the tester 3 Align the fou...
Страница 8: ...to 1000 ppm 0 10 to 10 00 ppt TDS Factor 0 5 Resolution 0 1 ppm 1 ppm 0 01 ppt Relative accuracy 1 full scale Calibration points Up to 3 points TDS Factor 0 40 to 1 00 Selectable Salinity Salinity ra...
Страница 9: ...TSPIN Conductivity TDS salinity pocket tester with batteries ELITECTSCUP Conductivity TDS salinity pocket tester with batteries Conductivity Standard Solutions and Sachets ECCON100BT 100 S cm KCl 480...