Connecting the EASY-Spray Column Plumbing
Examples of LC Connections to the EASY-Spray Column
Thermo Scientific
EASY-Spray Series Ion Source User Guide
27
Configuring the EASY-nLC 1200 Plumbing for Two Columns
Use the touch screen on the EASY-nLC 1200 to configure the use of two columns.
To configure the EASY-nLC 1200 for use with two columns
1. On the LC touch screen, press
Maintenance > Devices
.
2. In the Devices list, press
EASY-nLC (HPLC)
.
3. Press
Properties
, clear the
One Column Setup
check box, and then press
Apply
.
The Home page shows the two-column setup (
). For additional instructions,
refer to the LC instrument manual.
Figure 17.
EASY-nLC 1200 touch screen showing a two-column configuration
Connecting an RSLCnano System to the EASY-Spray Column
Follow the appropriate procedure for a one- or two-column setup for an RSLCnano system.
The two-column setup includes a precolumn connected across the autosampler valve.
•
To plumb a one-column setup for the RSLCnano instrument
•
To plumb a two-column setup for the RSLCnano instrument
Two columns
Note
To connect the RSLCnano system to the EASY-Spray column, you must use the
Thermo Scientific Dionex UltiMate 3000 RSLCnano EASY-Spray Application Kit
(P/N 6720.0395).