Connecting the EASY-Spray Column Plumbing
Examples of LC Connections to the EASY-Spray Column
28
EASY-Spray Series Ion Source User Guide
Thermo Scientific
To plumb a one-column setup for the RSLCnano instrument
1. Insert the nanoViper fitting on the 20 μm ID nanoViper capillary into port 5 of the
).
2. Insert the fitting on the other end of the capillary into the Viper union, tighten the fitting
by hand.
3. Connect the union to the EASY-Spray column to complete the one-column setup
).
Figure 18.
Example of a one-column setup for an RSLCnano system (direct injection)
CAUTION
Overtightening the fittings can irreparably damage their sealing surfaces.
For additional information about nanoViper fittings, see
EASY-Spray
column
Autosampler
Viper union
20 μm ID × 65 cm
20 μm ID
1 μL loop