Service Procedures
Doc. 031880-04 1/12
69
10.
Remove the 2-pin connector from the free end of the
yellow-and-black TTL cable. Reverse the positions of the two wires
and reattach them to the connector.
NOTE Use a screwdriver to loosen the locking screws and
remove the wires. Insert the wires in the reversed
position, and tighten the screws.
11.
Route the cable to the DX-120 rear panel and connect it to the
TTL-2 OUT
(see Figure 62). Double-check the connection: Make sure
the yellow wire is connected to
TTL-2 OUT
and the black wire is
connected to
Ground
.
12. Go on to Section B.4.
B.4
Power Connection
1. Connect the power cord (IEC 320 C13) from the main power
receptacle on the rear panel to a grounded power source.
2. Turn on the power to the Dionex RFC and to all other system
modules.
When attaching wires to the connector plug, be careful not to allow
stray strands of wire to short to the adjoining position on the
connector.
SHOCK HAZARD—To avoid electrical shock, use a grounded receptacle.
Do not operate the Dionex RFC or connect it to AC power mains without
an earthed ground connection.
The power supply cord is used as the main disconnect device. Make
sure the socket-outlet is located near the Dionex RFC and is easily
accessible.
Operation at AC input levels outside the specified operating voltage
range may damage the Dionex RFC.