4
–
Example Applications
Thermo Scientific
Product Manual for IonPac AS31 Columns
Page 20 of 45
155007-01
For Research Use Only. Not for use in diagnostic procedures.
4.3
Effect of Temperature on the Dionex IonPac AS31 Selectivity
The following chromatograms demonstrate the effect of temperature on the Dionex IonPac AS31 selectivity. Notice
there is a greater effect on the divalent anions (sulfate and thiosulfate) relative to monovalent anions (fluoride, chloride,
bromide, and nitrate) when temperature is changed from 15ºC to 30ºC.
Figure 2
Effect of Temperature on Dionex IonPac AS31 Selectivity
Column:
Dionex IonPac AG31/AS31, 2mm
Eluent:
24 mM KOH
Eluent Source:
Dionex EGC 500 KOH cartridge
Flow Rate:
0.30 mL/min
Inj. Volume:
2.5 µL
Temperature:
See Chromatogram
Detection:
Suppressed Conductivity,
Dionex ADRS 600 (2 mm)
AutoSuppression, recycle mode
Peaks (Standard):
mg/L
1.
Fluoride
5.0
2.
Sulfate
10.0
3.
Chloride
10.0
4.
Bromide
20.0
5.
Thiosulfate
20.0
6.
Nitrate
20.0
0
5
10
15
20
25
30
35
-1
10
1
3
2
4
5
Minutes
µS
-1
10
1
2
3
4
5
6
µS
15˚C
30˚C
6