Installation
Thermo Scientific
Product Manual for IonPac AS31 Columns
Page 10 of 45
155007-01
For Research Use Only. Not for use in diagnostic procedures.
2.
Installation
Read the instrument manuals. This manual assumes that you are using Thermo Scientific
Dionex instrumentation and are familiar with the installation and operation of the Thermo
Scientific Dionex Ion Chromatograph (IC). If you do not understand the operation of the
system, take the time to familiarize yourself with the various system components before
beginning an analysis.
The proper configuration of an Ion Chromatography System (ICS) is dependent on the column
format. Although they can typically use the same system, the use of a 2 mm or 4 mm column
requires different set up to ensure optimum performance. The selected format and analysis type
will affect the type of pump recommended. A gradient pump is designed to blend and pump
isocratic, linear, or gradient mixtures of up to four mobile phase components at precisely
controlled flow rates. An isocratic pump is for applications not requiring gradient or multi-eluent
proportioning capabilities. For high pressure applications (over 3000 psi), the use of high-pressure
consumables, components and fittings is required.
2.1
Thermo Scientific Dionex High Pressure Ion Chromatography Systems
A minimum of a Dionex High Pressure Ion Chromatography (HPIC) System is recommended
when running Dionex IonPac AS31 columns due to the higher backpressures generated at typical
operational flow rates. Systems should have the capability to operate up to at least 5000 psi.
Standard IC systems, with an upper limit of 3000 psi, are insufficient for proper column operation.
All systems should allow 100% metal-free operation to prevent column damage. This includes
pump heads and all flow paths.
Care should always be taken not to exceed the maximum operating pressure of the system
components. IC systems with lower backpressure capabilities are not recommended as
reduced flow rates may result in loss of performance.
Contact your local representative for information on how to customize your system to
your application needs.
2.2
System Requirements
The Dionex IonPac Columns are designed to run on Dionex Ion Chromatographs equipped with
suppressed conductivity detection. We recommend the use of ferrules and fittings rated with a
pressure of >5000 psi. The use of Thermo Scientific Dionex IC PEEK Viper
TM
fittings is
recommended to achieve consistent low dead volume connections and ensure optimum
chromatographic performance. Dionex IC PEEK Viper fittings are available in convenient kits
for Dionex ICS-5000
+
and ICS-6000 systems with conductivity detectors (Item # 088803).