Maintenance
|
Chapter 13
Changing the ultrapure cartridge
Note
Replace the ultrapure cartridge when the “Lim. val.pure w.“ message is displayed indicating
the purity dropped below acceptable levels or when the “New filter set” message is shown in the
display.
Step
Action
Figure
1
Switch the system off.
2
Remove the cover in front of the ultrapure cartridge and press the
snapper and pull the cover towards the front to remove it.
3
Press on the two quick-connectors at the inlet and outlet of the
ultrapure cartridge and remove the used cartridge.
4
We recommend performing a disinfection when an existing ultra-
pure cartridge is replaced.
5
Place the ultrapure cartridge in the rear section of the unit and
insert the two quick-connectors into the ultrapure cartridge. When
you hear an audible click you can be sure that the quick-action
fasteners have been inserted correctly.
6
Put the cover for the ultrapure cartridge back in place and switch
the unit on.
Snapper
Cover
Note
Refer to section “Disinfection” on page 82
for an explanation of how to perform disinfection
Note
The quick-connectors are attached to the unit in such a
manner so as to prevent any confusion (switching).
Ultrapure cartridge
Содержание dionex ic pure water
Страница 10: ......
Страница 14: ... Transport and packaging 4 Dionex IC Pure system Thermo Scientific Chapter 1 ...
Страница 18: ... Safety precautions 8 Dionex IC Pure system Thermo Scientific Chapter 2 ...
Страница 22: ... Intended use of the device 12 IC Pure Thermo Scientific Chapter 4 ...
Страница 28: ... Technical specifications 18 Dionex IC Pure system Thermo Scientific Chapter 5 ...
Страница 62: ... Installation 52 Dionex IC Pure system Thermo Scientific Chapter 7 ...
Страница 70: ... Initial start up 60 Dionex IC Pure system Thermo Scientific Chapter 10 ...
Страница 74: ... Operating elements 64 IC Pure Thermo Scientific Chapter ...
Страница 108: ... Waste disposal 98 Dionex IC Pure system Thermo Scientific Chapter 14 ...
Страница 112: ... Troubleshooting 102 Dionex IC Pure system Thermo Scientific Chapter 15 ...
Страница 116: ... Dionex IC Pure system spare parts 106 Dionex IC Pure system Thermo Scientific Chapter 16 ...
Страница 118: ... Consumables 108 Dionex IC Pure system Thermo Scientific Chapter 17 ...
Страница 119: ...Thermo Scientific Dionex IC Pure system 109 Terminal assignment ...
Страница 120: ... Terminal assignment 110 Dionex IC Pure system Thermo Scientific Chapter 18 ...
Страница 122: ... Maintenance record 112 Dionex IC Pure system Thermo Scientific Chapter 19 ...
Страница 127: ...Thermo Scientific IC Pure 117 ...