1100°C Box Furnace
330985H01 |
6-51
Main Controller: 1x8 & 5x16 Segment Programmable |
Chapter 6
4. To set the ramp unit, press SCROLL
button to select
‘RAMP.U’
and then
press DOWN
or UP
button to select
hour, min or sec
(In this example
the ramp unit is set in min).
5. To set the Dwell unit, press SCROLL
button to select
‘DWEL.U’
and then
Press
or
to select
hour or min
(In this example the dwell unit is set in
min).
6. To set the first Target Set point, press SCROLL
button to select
‘TSP.1’
. Then
Press DOWN
or UP
button to set the value.
7. To set the first Ramp rate, press SCROLL
button to select
‘RMP.1’
. Press
DOWN
or UP
button to set the value.
8. To set the first Dwell, press SCROLL
button to select
‘DWEL.1’
. Press DOWN
or UP
button to set the value.
9. Now repeat the above three steps (6, 7 & 8) to set remaining all segments.
NOTE
If not all the segments are used for a program, the Ramp & Dwell of each of the
subsequent Segments should be set to OFF.
A program will end in one of two ways, either revert to the control Set Point or
dwell at the temperature set in the last segment used. When a program finishes
on a dwell and the dwell time expires the temperature will revert to the control set
point.
Before running a program ensure that the control set point is set to Zero to avoid
unexpected heating at the end of the program.
When a Holdback is set, each segment used must have a Ramp Rate assigned to
it, in order for it to be recognized by the program.
Содержание BF51728
Страница 11: ...Safety Notes Chapter 1 1 10 330985H01 1100 C Box Furnace...
Страница 15: ...Introduction 2 14 330985H01 1100 C Box Furnace Chapter 2...
Страница 27: ...Start Up 5 26 330985H01 1100 C Box Furnace Chapter 5...
Страница 63: ...Maintenance 9 62 330985H01 1100 C Box Furnace Chapter 9...
Страница 69: ...Wiring Diagram 10 68 330985H01 1100 C Box Furnace Chapter 10...
Страница 81: ...Find out more at thermofisher com...