| Main Controller: 1x8 & 5x16 Segment Programmable
6-40
|
330985H01
1100°C Box Furnace
Chapter 6
The 8-segment programmable 3216c controller consists of microprocessor based
three-mode PID (Proportional, Integral and Derivative) and appropriate output switching
devices to control the furnace. The programmable controller can be used as a single set
point controller or as a programmable controller. The 8-segment digital model enables
eight segments of programming. The controller is capable of varying temperature or
process value with time through programming. A program is stored as a series of
segments and can be run once, repeated a set number of times or run continuously.
Some of the commonly used parameters in operator Level 1 and Level 2 of 3216c
controller are:
Table 2
Parameter Description and Accessibility in 3216c
No. Parameter
Description
Level
Access
Value
Page
No.
1
T.STAT
Timer Status
Level 1 + 2
Read/Write
rES
6-39
2
T.REMN
Timer Remaining
Level 1 + 2
Read Only
-
6-32
3
T.ELAP
Elapsed Time
Level 1 + 2
Read Only
-
6-33
4
A1.DHI
Deviation High Alarm Set Point
Level 1
Read/Write
50
6-31
5
A2.HI
High Temperature Alarm Set Point
Level 2
Read Only
1125
6-31
6
SP.SEL
Set point Select
Level 1 + 2
Read/Write
SP1
6-23
7
SP1
Set point 1
Level 1 + 2
Read/Write
0-1100
6-24
8
SP2
Set point 2
Level 1 + 2
Read/Write
0-1100
6-24
9
RAMPU
Set point Ramp Units
Level 1 + 2
Read/Write
Mins
6-24
10
SP.RAT
Set point Rate Limit
Level 1 + 2
Read/Write
Off
6-25
11
A.TUNE
Auto Tune Enable
Level 1 + 2
Read/Write
Off
6-26
12
UNITS
Display Units
Level 1 + 2
Read/Write
Deg C
6-26
13
TM.CFG
Timer Configuration
Level 1 + 2
Read/Write
Prog(10)
6-38
14
SS.SP
Soft Start Setpoint
Level 1 + 2* Read Only*
-
6-35
15
SS.PWR
Soft Start Power Limit
Level 1 + 2* Read Only*
-
6-35
16
T.T
Requested Time Duration
Level 1 + 2* Read Only*
-
6-35
17
THRES
Timer Start Threshold
Level 1 + 2
Read/Write
1
6-36
18
END.T
Timer End Type
Level 1 + 2
Read/Write
Dwell
6-37
19
Timer.TimeRes
Timer resolution
Level 1
Read Only
Mins
6-36
20
Level 2
Read/Write
6-36
21
TSP.1 to TSP.4
Target Set point 1 to Target Set point 4 Level 2
Read/Write
550
6-38
22
RMP.1 to RMP.4
Ramp Rate 1 to Ramp Rate 4
Level 2
Read/Write
OFF
6-38
23
DWEL.1 to DWEL.4 Dwell Time 1 to Dwell Time 4
Level 2
Read/Write
1 hour
6-39
24
PB
Proportional Band
Level 2
Read/Write
15
-
25
TI
Integral Time
Level 2
Read/Write
95
-
26
TD
Derivative Time
Level 2
Read/Write
16
-
27
LBT
Loop Break Time
Level 2
Read/Write
30 mins
6-31
28
PV.OFS
PV Offset
Level 2
Read/Write
0
6-30
Содержание BF51728
Страница 11: ...Safety Notes Chapter 1 1 10 330985H01 1100 C Box Furnace...
Страница 15: ...Introduction 2 14 330985H01 1100 C Box Furnace Chapter 2...
Страница 27: ...Start Up 5 26 330985H01 1100 C Box Furnace Chapter 5...
Страница 63: ...Maintenance 9 62 330985H01 1100 C Box Furnace Chapter 9...
Страница 69: ...Wiring Diagram 10 68 330985H01 1100 C Box Furnace Chapter 10...
Страница 81: ...Find out more at thermofisher com...