![Thermo Scientific Barnstead Smart2Pure 12 UF Скачать руководство пользователя страница 90](http://html1.mh-extra.com/html/thermo-scientific/barnstead-smart2pure-12-uf/barnstead-smart2pure-12-uf_operating-instructions-manual_1105414090.webp)
15
Trouble shooting
80
Smart2Pure
Thermo Scientific
Water leak
• Leaky tubing connection
• Feedwater pressure > 6 bar
• Component defect (e.g. UV
quartz tube)
• Check and seal the tubing con-
nection
• Install a pressure reducer
• Contact the Thermo Fisher
Scientific service department
Dispensed water flow rate is too
low
• Ultrafilter is blocked
• 0.2 micron final filter is
blocked
• Air in 0.2 micron final filter
• Replace ultrafilter (see
“Changing
the ultrafilter”
on
page 71
)
• Replace the 0.2 micron final
filter
• Vent the 0.2 micron final filter
Error message:
Red LED in second line is light
up “Limit value Cond.“
• exhausted ultrapure cartridge
• Limiting value set too low
• Replace with new ultrapure
cartridge
• Check and reset the limiting
value
Error message:
Red LED in second line is light
up “Limit value Temp.“
• Temperature in system is too
high
• Limiting value is set too low
• Feedwater temperature is too
high
• Reduce temperature by letting
water flow out
• Check and reset the limiting
value
• Reduce the feedwater tempera-
ture
Error message:
'
Red LED in second line is light
up “Status Sens.“
• Interruption in cable to con-
ductivity measuring cell
• Measuring cell defect
• Check the cable and plug con-
nection
• Contact Thermo Scientific for
replacing
Error message:
Red LED in second line is light
up “Status Sens.“
The display shows “- -“ with
audible alarm.
• Interruption in cable to con-
ductivity measuring cell
• Temperature sensor in mea-
suring cell defect
• Check the cable and plug con-
nector
• Contact Thermo Scientific for
replacing
Error
Cause
Remedy
Содержание Barnstead Smart2Pure 12 UF
Страница 10: ...Contents viii Smart2Pure Thermo Scientific ...
Страница 14: ...1 Transport and packaging 4 Smart2Pure Thermo Scientific ...
Страница 18: ...2 Safety Precautions 8 Smart2Pure Thermo Scientific ...
Страница 24: ...3 Extend of delivery 14 Smart2Pure Thermo Scientific ...
Страница 26: ...4 Intended Use of the device 16 Smart2Pure Thermo Scientific ...
Страница 56: ...7 Installation Connections of the Smart2Pure 12 system 46 Smart2Pure Thermo Scientific ...
Страница 58: ...8 Flow chart 48 Smart2Pure Thermo Scientific ...
Страница 60: ...9 How the Smart2Pure system functions 50 Smart2Pure Thermo Scientific ...
Страница 66: ...11 Operating Elements 56 Smart2Pure Thermo Scientific ...
Страница 88: ...14 Waste disposal 78 Smart2Pure Thermo Scientific ...
Страница 98: ...17 Consumables 88 Smart2Pure Thermo Scientific ...
Страница 99: ...Thermo Scientific Smart2Pure 89 18 Terminal assignment ...
Страница 100: ...18 Terminal assignment 90 Smart2Pure Thermo Scientific ...
Страница 102: ...19 Maintenance record 92 Smart2Pure Thermo Scientific ...
Страница 107: ...Thermo Scientific Smart2Pure 97 I ...
Страница 108: ...Index ...
Страница 109: ...Thermo Scientific Smart2Pure 99 I Index ...