28
Resetting the UV Timer
When the Nanopure TOC - UV is in an (Idle) position:
1.
From the (Idle) mode, when display reads,
“Nanopure (Idle) xx:xx:xx xx/xx/xx,” press the
UP or DOWN arrow until the display reads,
“Reset Timers?” Press ENTER.
2.
Press the UP or DOWN arrow until the display
reads, “Reset UV Timer.”
3.
Press the UP or DOWN arrow to select “Yes.”
4.
Press ENTER and the timer will be reset for
approximately five months of operational time.
The display will now display “Rest Cleaning Timer.” Press
BACK two times to return to the (Idle) mode.
Resetting the TOC Lamp Timer
When the Nanopure TOC - UV is in an (Idle) position:
1.
From the (Idle) mode, when display reads,
“Nanopure (Idle) xx:xx:xx xx/xx/xx,” press the
UP or DOWN arrow until the display reads,
“Reset Timers?” Press ENTER.
2.
Press ENTER until the display reads, “Reset
TOC Timer.” It will also read, on the right side of
the second line, the hours of usage which have
occurred since the TOC Timer was last reset.
3.
Press the UP or DOWN arrow to select “Yes.”
4.
Press ENTER and the timer will be reset for
1820 hours of operational time.
The display will now display “Reset TOC Timer.” Press
BACK two times to return to the (Idle) mode.
U
SER
S
ETTINGS
Note
When the user begins receiving
“Change UV Lamp” reminders, it will
be time to change UV lamp and
reset UV timer as described in
“Resetting the UV Timer.”