Tubing Installation
To make tubing connections:
1.
Make sure the tubing is cut off reasonably
square and that no plastic burrs or ridges
are present.
2.
*Mark from end of tube the length of insertion.
Tube size
Insertion length
1/4 O.D.
1.75 cm (11/16”)
3/8 O.D.
1.9 cm (3/4”)
3.
Wet the tube end with water and insert tube
straight into fitting until it bottoms out on interior
shoulder and insertion mark is no longer visible.
4.
To remove tubing, push collet toward fitting body
and pull on tubing. To reuse fitting, begin assembly
at step one.
Electrical Connections
The Nanopure comes with a 120 V power cord with (2) 1.6
amp fuses and 240 V power cords with (2) 0.63 amp fuses.
The Nanopure is not shipped with fuses installed in the fuse
draw of the power module. Before connecting the power
cord to the power module, install the proper fuses in the
fuse holder. Refer to the figure on the left.
I
NSTALLATION
Fuse Holder
Power Switch
Power Cord
Figure 7: Electrical Connections
COLLET
TUBING
**
TYPICAL
FITTING
Figure 5: Tubing Adapter Installation
14
Figure 6: Tubing removal
Содержание Barnstead Nanopure TOC - UV