4-44 AutoXP Users Guide
Thermo Fisher Scientific
HART
7. User enters actual measured value.
Figure 4-81. Enter meter value.
8. AutoXP will modify the output to provide 20mA.
Figure 4-82. Fld dev output equal to reference meter
If the output goes to 20mA, proceed to step 9 below.
If the output is not 20mA, go back to step 7.