Thermo Fisher Scientific
4-17 AutoXP Users Guide
HART
7.
If a 1 point calibration
i
s required
,
go to step 17 below.
8.
If a 2 point calibration
i
s required
,
go to step 13 below.
For a 3 point calibration,
su
pply a mid-range value to the input.
Figure 4-29. Supply MID degF to RTD Input
9. DD will display live updates of actual input value.
10. Press ENTER/OK when value is stable.
Figure 4-30. EU Value
11.
Enter actual Mid value
su
pplied to the input.
Figure 4-31. Enter actual MID value
supplied
to the input