Antaris II and Antaris MX Site and Safety Information
55
Site Preparation Checklist
Use this check list to ensure that all site-preparation steps have been
performed. You will then be ready to schedule the installation.
General
Affected administrators, such as MIS, shipping, maintenance,
metrology, users, and regulatory employees, have been informed that
the system is on order and when it will be installed.
The system remains packed until the Thermo Fisher Scientific service
engineer arrives.
The system administrator is scheduled to be present during the
installation to complete the following tasks:
- Set up users in the Windows operating system
- Set up Windows security features
- Change settings for the R
ESULT
Operation audit log, if necessary
- Set up another R
ESULT
administrator account, if necessary
Note
A system administrator must be present during installation of an Antaris
system. The Thermo Fisher Scientific service engineer may need to delay or
reschedule the installation if a system administrator is not available.
Environmental considerations
The temperature stays relatively constant between 15 °C and 35 °C
(59 °F and 95 °F); between 20 °C and 22 °C (68 °F and 72 °F) for best
performance.
The humidity (non-condensing) stays between 5 % and 95 %;
between 20 % and 80 % for best performance.