Specifications
Coolant gas
Some rackmount enclosure require a flow of coolant gas. The following
specifications can be used to ensure that an unpurged instrument is
protected.
Item
Specification
Source line pressure
20 psig min
1.4 atm min
150 psig max
10.2 atm max
Purge pressure
10 psig min
0.7 atm min
20 psig max
1.4 atm max
Flow rate
15 scfh min
7.1 l/m min
20 scfh max
9.4 l/m max
Filtering 10-micron
filter
removes
particulate
matter and oil
Dew point
-70 °C
(-94 °F)
or below
for best
performance
Purge gas generator
The following illustrations show the dimensions of the purge gas
generators and the locations of holes that can be used for mounting the
generators on a wall. Models 75-45 and 75-45EU fit easily on top of a
bench or table. When not wall-mounted, Models 75-52 and 75-52EU
should be placed on the floor.