![Thermo Scientific ALPHA DO 2000 PG POLAROGRAPHIC DISSOLVED OXYGEN CONTROLLER TRANSMITTER Скачать руководство пользователя страница 42](http://html1.mh-extra.com/html/thermo-scientific/alpha-do-2000-pg-polarographic-dissolved-oxygen-controller-transmitter/alpha-do-2000-pg-polarographic-dissolved-oxygen-controller-transmitter_manual_1104618042.webp)
Instruction manual
Alpha
DO 2000
8. Technical Specifications
8.1 General Specifications
Dissolved Oxygen
Measuring range
0.00 to 20.00 mg/l or ppm
0.0 to 200.0 %
Resolution
0.01 mg/l or ppm or
0.1 %
Relative accuracy
±1 % of full-scale reading
Temperature
Measuring range
-9.9 to +125.0
o
C or
+15.0 to +257.0
o
F
Resolution 0.1
o
C or
0.1
o
F
Relative accuracy
±0.5
o
C or
±1.0
o
F
Sensor
NTC 22 k
Ω
thermistor
Compensation
Temperature Compensation
Automatic or Manual
Salinity Compensation
0.0 to 50.0 ppt
(Manual setting and automatic correction)
Pressure Compensation
0.740 to 3.000 bar
555 to 2250 mmHg
10.73 to 43.51 PSI
(Manual setting and automatic correction)
Set-point and Control Function
Function Switchable
Off
Limit Control
P/PI Control (Pulse length / Pulse
frequency)
Integral Action Time (IAT)
0.0 to 999.9 minutes
Adjustable period with pulse length
controller
0.5 to 20 seconds
Adjustable period with pulse
frequency controller
60 to 120 pulses/min
Pickup / Dropout delay
0 to 1999 seconds
Wash cycle
0.1 to 199.9 hours
Wash duration
1 to 1999 seconds
Switching DO hysteresis
0.1 to 1.0 mg/l or 1.0 to 10.0 %
Contact outputs, controller
1 SPDT, 3 SPST relays
Switching voltage
max. 250 VAC
Switching current
max. 3A
Switching power
max. 600 VA
34
Содержание ALPHA DO 2000 PG POLAROGRAPHIC DISSOLVED OXYGEN CONTROLLER TRANSMITTER
Страница 1: ...Alpha DO 2000 PPG Processor Transmitter Dissolved Oxygen ppm m L...
Страница 4: ......
Страница 6: ......
Страница 13: ...Instruction manual Alpha DO 2000 2 3 Overview Wall mounting version Panel mounting version 5...
Страница 54: ......
Страница 55: ......