![Thermo Scientific ALPHA DO 2000 PG POLAROGRAPHIC DISSOLVED OXYGEN CONTROLLER TRANSMITTER Скачать руководство пользователя страница 27](http://html1.mh-extra.com/html/thermo-scientific/alpha-do-2000-pg-polarographic-dissolved-oxygen-controller-transmitter/alpha-do-2000-pg-polarographic-dissolved-oxygen-controller-transmitter_manual_1104618027.webp)
Instruction manual
Alpha
DO 2000
6. Setup Mode
6.1 Enter Setup Mode
In the Setup mode the transmitter can be configured to your individual requirements.
MEAS
SETUP
SETUP
SETUP
SETUP
SETUP
SE TUP
SETUP
SETUP
SETUP
HOLD
HOLD
HOLD
HOLD
HOLD
HOLD
HOLD
HOLD
HOLD
%
°C
Temperature settings
see section 6.3
Relay A (set point 1) settings
see section 6.4
Relay B (set point 2) settings
see section 6.4
Controller settings
see section 6.5
Current output 1 settings
see section 6.6
Current output 2 settings
see section 6.7
Wash relay settings
see section 6.8
Calibration
see section 6.10
SETUP
HOLD
Configuration settings
see section 6.9
OFS
13.0
DO
Offset settings
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
see section 6.2
CAL
1. While in Measurement mode
press ENT key
.
2. The display prompts you to enter the security code.
Set the security code with
▲
or
▼
key
to:
−
“SCD 22” if you want to change parameter
settings
−
“SCD 000” if you only want to view parameter
settings (
view only mode
)
3.
Press ENT key
.
Î
NOTE:
To exit Setup mode at any time
press
▲
and
▼
keys simultaneously
(escape). The transmitter returns to the Measurement mode.
19
Содержание ALPHA DO 2000 PG POLAROGRAPHIC DISSOLVED OXYGEN CONTROLLER TRANSMITTER
Страница 1: ...Alpha DO 2000 PPG Processor Transmitter Dissolved Oxygen ppm m L...
Страница 4: ......
Страница 6: ......
Страница 13: ...Instruction manual Alpha DO 2000 2 3 Overview Wall mounting version Panel mounting version 5...
Страница 54: ......
Страница 55: ......