![Thermo Scientific ALPHA DO 2000 PG POLAROGRAPHIC DISSOLVED OXYGEN CONTROLLER TRANSMITTER Скачать руководство пользователя страница 36](http://html1.mh-extra.com/html/thermo-scientific/alpha-do-2000-pg-polarographic-dissolved-oxygen-controller-transmitter/alpha-do-2000-pg-polarographic-dissolved-oxygen-controller-transmitter_manual_1104618036.webp)
Instruction manual
Alpha
DO 2000
6.7 Current Output 2 Settings (CUR 2)
In this, you set the current output range of the transmitter for temperature
values.
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
SETUP
SETUP
HOLD
HOLD
°C
°C
°C
°C
SETUP
SETUP
HOLD
HOLD
ENT
ENT
ENT
ENT
ENT
ENT
ENT
1
2
2
3
3
4
4
1. From the setup mode, select the “CUR 2” screen and then
press ENT key
.
2.
Selecting the output type: Press
▲
or
▼
key
to select the desired output
type: 0-20 mA or 4-20 mA.
Press ENT key
to confirm your selection.
3.
Setting the temperature value at which the transmitter output will be 4
mA (or 0 mA): Press
▲
or
▼
key
to set the temperature value (setting range:
-9.9 to 115°C or 15 to 237 °F) to be equivalent to 4 mA (or 0 mA).
Press ENT
key
to confirm your setting.
4.
Setting the temperature value at which the transmitter output will be 20
mA: Press
▲
or
▼
key
to set the temperature value (setting range: -0.1 to
125°C or 35 to 257 °F) to be equivalent to 20 mA.
Press ENT key
to confirm
your setting.
The controller goes back to “CUR 2” screen.
Continue with rest of the Setup procedures by pressing
▲
or
▼
key or return
to the Measurement mode by pressing
▲
and
▼
keys (escape)
simultaneously.
28
Содержание ALPHA DO 2000 PG POLAROGRAPHIC DISSOLVED OXYGEN CONTROLLER TRANSMITTER
Страница 1: ...Alpha DO 2000 PPG Processor Transmitter Dissolved Oxygen ppm m L...
Страница 4: ......
Страница 6: ......
Страница 13: ...Instruction manual Alpha DO 2000 2 3 Overview Wall mounting version Panel mounting version 5...
Страница 54: ......
Страница 55: ......