3
Installation
Installation of the AI 3000/AS 3000 II on the GC
Thermo Scientific
AI 3000/AS 3000 Sampling System - Operating Manual
49
I
nstallation of the AI 3000/AS 3000 II on the GC
To install the AI 3000/AS 3000 II on the TRACE 1300/TRACE 1310, TRACE GC Ultra,
and FOCUS GC gas chromatographs, please refer to the instructions reported below:
The procedure includes the following steps:
•
Installation of the Sampling Unit
•
•
•
Installation of the Sampling Unit
Proceed as follows:
1. Lift the sampling unit, see
, and insert it into the guide pivot located
on the sampling system support. See
.
Figure 42.
Installation of the Sampling Unit
2. Insert the proper centering plate for TRACE GC Ultra or TRACE 1300/TRACE 1310,
see
, into its seat of the sampling unit base paying attention that the guide hole,
present on the arm of the centering plate, correctly fits the injector nut.
Figure 43.
Centering Plates
Sampling Unit
Guide Pivot
Centering Plate for
TRACE GC Ultra
Centering Plate for
TRACE 1300/1310
PTV
SSL
Содержание AI 3000