Xcalibur Diagnostics for the PDA Detector
Calibrating the PDA Detector from the Xcalibur Data System
Thermo Scientific
Accela PDA Detector Hardware Manual
55
Figure 48.
Page 8 of the Wavelength Calibration wizard
11. Check that the delta values are within ±1 nm, and then do one of the following:
• If the Delta values are acceptable, proceed to
• If the delta values are not within the range of ±1 nm, do not export the results, and
proceed to
. Complete the calibration procedure, and then repeat the
wavelength calibration. If, after applying a second calibration, the Delta values are
still not within the range of ±1 nm, call your Thermo Fisher Scientific service
representative for assistance.
12. (Optional) Print a report of the calibration results:
a. Click
Export Results
to print the results to a file.
The Save As dialog box appears (see
).