Xcalibur Diagnostics for the PDA Detector
Calibrating the PDA Detector from the Xcalibur Data System
Thermo Scientific
Accela PDA Detector Hardware Manual
45
Figure 33.
Calibration page
4. To start the Dark Current Calibration wizard, in the Dark Current Calibration area click
Calibrate
.
The preconditions page appears (see
).
Figure 34.
Preconditions page
5. Read the preconditions and verify that they have been met:
• If the preconditions have been met, click
Next
to proceed with the calibration.
• If the preconditions have not been met, click
Cancel
. Then prepare the Accela PDA
Detector for calibration and begin this procedure again.
6. On page 2 of the Dark Current Calibration wizard, observe the status readback as the
calibration proceeds (see