Single running interface
Press “STOP”, and the
interface of asking you
whether to stop the program
will appear.
Press “YES” to stop the program and to go
back to the main interface.
Press “NO” to go back to the single running
interface.
Press “CAL” to calibrate the temperature
adjustment interface. (If the external sensor is
being used, then there would be no “CAL.”,
and the temperature is unable to be adjusted.)
Input the correct temperature into the box that pops out.
Caution
a) To ensure the accuracy of temperature, the instrument should
be calibrated after temperature stabilizes for 30 minutes.
b) Calibrate the instrument with the qualified Grade II standard filled
thermometer
c) Calibration point: The middle hole of the heat block. Pour paraffin
oil into the well, and soak the thermometer bulb.
Instruction: During the operation of the instrument, the temperature and
the time should not be changed. If it they should, stop the program and
enter into the setting interface to change them.
Listed files interface
Press “BACK” to go back to the main interface.
Enter into the setting
interface of the specific file in
accordance with its name. If
the file has a password, then
enter into the password
inputting interface.
6-2
Drybath Incubator
Thermo Scientific
Section 6
Operation Guidance
Operation Instructions
(continued)