8

     Model 60

i

 Quick Start Guide 

Thermo Fisher Scientific 

Rear Panel 
Connector Pin 

Terminal Block 
Screw Number 

Signal Description 

5 17 Analog_In4 

6 18 Analog_In5 

7 19 Analog_In6 

8 20 Ground 

9 21 Analog_In7 

10 22 Analog_In8 

11 23 Ground 

14 24 Ground 

 

Before the power cord is connected, be sure that the front panel power 
switch is OFF and be sure to verify that the local power matches the label 
on the back of the analyzer. 

WARNING  

The Model 60

i

 is supplied with a three-wire grounded power 

cord. Under no circumstances should this grounding system be defeated. 

  

Use the following procedure when starting the instrument. 

1.

 

Turn the power ON and allow two hours for the instrument to 
stabilize. 

2.

 

Set instrument parameters such as operating ranges and averaging times 
to appropriate settings. For more information about instrument 
parameters, see the “Operation” chapter in the “Model 60

i

 Instruction 

Manual.” 

3.

 

Before beginning the actual monitoring, perform a calibration as 
described in the “Calibration” chapter in the “Model 60

i

 Instruction 

Manual.” 

 

Power Connections 

 

Startup 

Содержание 60i

Страница 1: ...Model 60i Quick Start Guide Multi Gas Analyzer Part Number 106474 00 23Jul2009 Example of a Direct Extractive Continuous Emissions Monitoring System CEMS ...

Страница 2: ...ep Action 1 Connect the sample line from the chiller bypass pump to the SAMPLE bulkhead on the rear panel Ensure that the sample line is not contaminated by dirty wet or incompatible materials All tubing should be constructed of FEP Teflon with an OD of 1 4 inch Note Sample gas must be filtered and dried when delivered to the analyzer The filter should have a nominal pore size of 1 micron or less ...

Страница 3: ...pan gases and zero air to CAL IN via inch tubing 4 Connect the EXHAUST bulkhead to a suitable vent The exhaust line should be inch OD with an ID of 1 8 inch or greater The length of the exhaust line should be less than 10 feet Verify that there is no restriction in this line Note If calibration gases will not be introduced through the sample probe see the Calibration chapter in the Model 60i Instr...

Страница 4: ...be Barrel Calibration gas connections to manifold All calibration gases are regulated to the same pressure Usually 10 to 30 psi depending on choice of capillary or needle valve used downstream Calibration Gas Manifold Capillaries or needle valve restrict flow to 1 5 to 2 0 LPM for direct calibration or 5 to 6 LPM for system calibration Heated Sample Line Calibration Line Heated Particulate Filter ...

Страница 5: ... analog outputs and the most commonly used digital inputs Refer to Table 3 for information on optional analog current outputs and optional analog voltage outputs I O expansion board 25 pin connector The following terminal block PCB assemblies are available for iSeries instruments I O terminal block PCB assembly 37 pin standard D O terminal block PCB assembly 37 pin standard 25 pin terminal block P...

Страница 6: ...7 2 4 6 8 10 12 Signal Grounds NA 1 13 Power Fail NC NA 20 14 Power Fail Com NA 21 15 Power Fail NO NA 3 16 TTL Input 1 Turn Pump Off 4 17 TTL Input 2 Trigger Blow Back Pin 1 Pin 20 Pin 3 Digital Input Channel 1 Pin 4 Digital Input Channel 2 Pin 23 Digital Input Channel 3 Pin 24 Digital Input Channel 4 Pin 6 Digital Input Channel 5 Pin 25 Digital Input Channel 6 Pin 7 Digital Input Channel 7 Pin 2...

Страница 7: ...e All channels are user definable If any customization has been made to the analog output configuration or TTL inputs the default selections may not apply Defaults shown are for a fully equipped analyzer Table 3 Optional Analog Current Outputs and Analog Voltage Inputs I O Expansion Board 25 Pin Connector Rear Panel Connector Pin Terminal Block Screw Number Signal Description 15 1 IOut1 13 2 Isola...

Страница 8: ...The Model 60i is supplied with a three wire grounded power cord Under no circumstances should this grounding system be defeated Use the following procedure when starting the instrument 1 Turn the power ON and allow two hours for the instrument to stabilize 2 Set instrument parameters such as operating ranges and averaging times to appropriate settings For more information about instrument paramete...

Страница 9: ...not expose the instrument to vibration or magnetic interference Do not remove adjust the small circular entrance or exit mirrors in the optical bench Use an external pump with atmospheric bypass dump to deliver the sample gas to the instrument Filter condition sample stream Sample dew point must be between 2 C and 8 C Operate the instrument in a climate controlled environment 15 30 C Set DHCP to O...

Страница 10: ...rights reserved Specifications terms and pricing are subject to change Not all products are available in all countries Please consult your local sales representative for details Thermo Fisher Scientific Air Quality Instruments 27 Forge Parkway Franklin MA 02038 1 508 520 0430 www thermo com aqi ...

Страница 11: ... describes possible alarm messages that may be displayed during start up and provides the recommended action to take to restore the system to normal operation Troubleshooting Alarm Messages Alarm Message Possible Cause Action Sample Flow Pump defective Pump disconnected Defective or disconnected flow switch cable Improper alarm threshold setting Plumbing connection configuration Repair or replace ...

Страница 12: ... Intensity Analog Input Readings Analog Input Voltages Digital Inputs Relay States Test Analog Outputs Instrument Config Contact Info Alarms Concentration Alarms Other Alarms System Alarms Service Absorbances Show User Zero Show User Span Pressure Calibration O Sensor Calibration Preamp Board Cal Temperature Calibration Analog Out Cal Lock Analog Outputs Analog Out Lock Type Set Analog Lock Percen...

Отзывы: