Thermo Fisher Scientific
Model 60
i
Quick Start Guide
3
System includes 61i
Calibrator?
Cap CAL IN and CAL
OUT bulkheads
Step 2
Disable internal
solenoid valves
N
Y
Step 3
Step 4
Step 4
Step 1
Step Action
2
Connect umbilical calibration gas line to CAL OUT.
3
Connect the output from the 61i calibration manifold (span gases and zero air) to CAL
IN via ¼-inch tubing.
4
Connect the EXHAUST bulkhead to a suitable vent. The exhaust line should be ¼-inch
OD with an ID of 1/8-inch or greater. The length of the exhaust line should be less than
10 feet. Verify that there is no restriction in this line.
Note
If calibration gases will not be introduced through the sample probe,
see the “Calibration” chapter in the “Model 60
i
Instruction Manual” for
alternative plumbing arrangement
.
▲
Figure 2.
Flowchart of Setup for Systems with a 61i Calibrator