10
Changing Chamber
Temperature Control Setpoint
1. The chamber temperature is continuously dis-
played, unless a key is pressed.
2. To display the setpoint temperature press the
SET key. The setpoint temperature will remain
on the display for 1 sec. after the SET key is
released.
3. To change the setpoint temperature, press and
hold SET while pressing the
L
or
M
keys to
change the setting. (Note that when either
L
or
M
keys are first pressed, the display will begin
to change slowly at first, then increase change
rate after a few seconds.)
4. Select the desired chamber setpoint then
release all keys. The controller display will
revert back to actual chamber temperature after
a few seconds. Allow up to an hour for unit to
stabilize at the new setpoint temperature.
After the incubator temperature has stabilized, place sam-
ples into incubator chamber. For best results, arrange
samples evenly throughout the chamber. Liquid samples
should be covered to prevent evaporation and eventual
frost build-up on evaporator coils, particularly when oper-
ating below ambient.
The Low Temperature Incubator has an internal 2 amp.
convenience outlet inside chamber to operate devices
such as: shakers, rotators, photosynthesis lights, record-
ing thermometers etc.
Accessing Controller Menus
The temperature controller has three menu selections that
can be accessed by pressing and holding the MENU key
for 3 sec. To access menu parameter, press and hold the
SET key. To change the parameter, use the
L
or
M
keys
CAL Menu
- Calibration Menu is used to change the tem-
perature offset value to correct for differences in chamber
O
PERATION