Model 3050/3052 5-1
Thermo Scientific
Section 5
Specifications
Exterior Dimensions
. .7.13”W x 5.13”H x 9.25 F-B
. . . . . . . . . . . . . .(18.1 cm x 13.1 cm x 23.5 cm)
Gas Connections
. . . . . .Inlet - two ¼” barb fittings;
. . . . . . . . . . . . . . . . .Outlet - one ¼” barb fitting
Inlet Pressure
. . . . . . . . . . . . . . . .15 PSIG, ±5 PSIG
Construction
. . . 18 Ga. CRS, Heavy-gauge metal w/powder coated finish
Electrical
. . . . . . . . . . . . .Switchable from 115/230V
100-115V, 1PH, 50/60 Hz, 0.3 FLA
200-230V, 1PH, 50/60 Hz, 0.15 FLA
Not to exceed ±10% of nominal
Power Connection
. . . . . . . . . .IEC 320 Female Plug
Remote Alarm Contact
. . RJ-11 Phone Jack (C, NO, NC)
Contact Rating
. . . . . . . . . . . . . . . . . .30V, 1A Max
Shipping Weight
. . . . . . . . . . . . . . . .13 lbs. (5.9 kg)
The Model 3050/3052 Gas Guard is designed to be electrically safe in the
following environmental conditions:
Indoor use only
Altitude: Up to 2,000 meters
Temperature: 5°C to 40°C
Humidity: 80% RH at or below 31°C, decreasing linearly to 50% RH
at 40°C
Mains Supply Fluctuations: ±10% of nominal voltage
Installation Category II
1
Pollution Degree 2
2
Class of Equipment
1 Installation category (overvoltage category) defines the level of transient overvoltage which the instrument
is designed to withstand safely. It depends on the nature of the electricity supply and its overvoltage
protection means. For example, in CAT II which is the category used for instruments in installations supplied
from a supply comparable to public mains such as hospital and research laboratories and most industrial
laboratories, the expected transient overvoltage is 2500V for a 230V supply and 1500V for a 120V supply.
2 Pollution Degree describes the amount of conductive pollution present in the operating environment.
Pollution Degree 2 assumes that normally only non-conductive pollution such as dust occurs with the
exception of occasional conductivity caused by condensation.