Reciprocal Shaking Water Baths
11-3
Thermo Scientific
Part Replacement
Section 11
Replace Power Supply
1. Remove connectors J101, J102, J103, and J105 from board. Use extra care when
removing J102 to not bend connector pins.
2. Remove five nuts from board and remove board.
3. Reverse above procedure to install new power supply board.
Replace Bearing Assembly
1. Remove bolt on top of plastic bearing and remove plastic bearing.
2. Disconnect tray from push rod and remove push rod.
3. Remove jam nut from control side of bearing assembly.
4. Slide bearing assembly out and into bath side of unit.
5. Remove bearing cap and wiper seal from assembly.
6. Reverse above procedure to install new bearing assembly.
Replace CPU or Display/
Keyboard Board. See note on
following page.**
1. Remove connectors J302 and J303 from CPU board.
2. Remove connector J102 from power supply board. Use extreme care when removing
J102 to not bend connector pins.
3. Remove 4 nuts from CPU board and remove CPU board and Keyboard/Display board.
4. Remove J301 from CPU board. Use extreme care when removing J301 to not bend
connector pins. Remove board.
5. Reverse above procedure to install new board.
6. The software must be configured to match the Bath model. To enter configuration
mode, press the upper left push-button while applying power. The upper left push-but-
ton is labeled "Select/Enter." This push-button is used to enter configuration codes and
steps through the different levels of configurations. When a triangle is shown in the
chart, the push-button is to be depressed once for each symbol.
(continued)