5

7.

OPERATION

CAUTION

EXERCISE CARE WHEN USING ACIDIC OR
CAUSTIC SOLUTIONS AS THEY WILL ATTACK
THE GALVANIZED STEEL BATH BODY IF SPILLED
INTO THE BATH. IF SPILLS DO OCCUR, THE BATH
LIQUID SHOULD BE IMMEDIATELY DRAINED AND
THE UNIT THOROUGHLY FLUSHED. SPILLS AND
C O N D E N S A T I O N   S H O U L D   B E   C L E A N E D /
REMOVED FROM ALL METAL SURFACES AFTER
EACH USE.

WARNING

EXPLOSION, IMPLOSION OR THE RELEASE OF
TOXIC OR FLAMMABLE GASES RISING FROM
M A T E R I A L   B E I N G   H E A T E D   I S   T H E   S O L E
RESPONSIBILITY OF THE USER.

7.01

When filling the bath with water, allowance

must be made for the displacement of water
upon immersion of samples. The maximum
liquid level should  be 1 inch below the top of the
pan. 

Do not  use deionized water.

7.02

To conserve energy, reduce evaporation

and increase temperature control accuracy, use
the gable cover provided. Do not use aluminum
foil as a cover; as it may cause corrosion due to
an electrochemical reaction.

1. Add distilled water  to bath. Position the

temperature probe tip 4 1/2" from the top
edge of pan.  Adjust O-ring if necessary.
Probe must always be immersed in the pan
during operation.  Water level should be at
least 1-1/2 inches above the top pump outlet
for adequate water circulation. The water
level should be approximately 1 to 1 1/2
inches from the top of the bath when the bath
is fully loaded.

2. Power up: Depress the power switch lo-

cated at the lower left hand corner of the
bath. The unit is fully operational upon appli-
cation of power and the display will read
actual bath temperature.

3. To Select the Operating Temperature:

Model 2860 has four user selectable
temperature setpoints of 35.0, 41.5, 44.5
and 45.5°C.  To select the desired operating
temperature, press the 

V

 or down 

W

 button

until the desired temperature is displayed.
Press Enter button to accept the new setpoint.
The display will flash, indicating that the new
setting has been entered.  When the flashing
stops, the display returns to actual bath
temperature.

4.  To Calibrate:

a. Calibration should be performed to match

a stable bath's actual temp display to a
calibrated thermometer.

b. Press CALIBRATE button once. Display

flashes to indicate Calibrate mode.

c. Press 

V

 or 

W

 button until the display

matches the readout on a certified ther-
mometer. Press ENTER to set new value.

d. If ENTER is not pressed, the new value will

not be set and display will return to the
previously displayed temperature.

e. Display should now match your calibrated

thermometer. The unit will stabilize
automatically to the set temperature.

 

Periodic Safety Check

Test the operation of the Hi-Limit thermostat, as
follows, every three months.
1. Locate the Hi-Limit thermostat adjustment

to the right of the power switch.

2. When delivered, the thermostat adjustment

is turned fully clockwise. In this position,
the Hi-limit light is OFF.

3. After the bath temperature has stabilized,

use a standard, flatblade screwdriver to
adjust the thermostat.

4. Turn the thermostat adjustment counter-

clockwise until the Hi-limit light is ON.

5. Then turn the thermostat adjustment

clockwise 1/8 of a turn at a time, until the
Hi-limit light goes OFF.

At this setting, if the bath temp increases
above the selected setting, the Hi-limit
activates.

Содержание 2860

Страница 1: ... Be Re quire d Pre ss or until d e sire d te mp e rature is d isplaye d Display will the n re turn to actual te mpe rature Pre ss ENTER Ne w Se t Point is acce pte d To Se le ct Te mp e rature Se le cte d Te mp e rature s Are Factory Prese t Pre ss ENTER calib ration is comp le te Pre ss or to match re ad out to the rmome te r Pre ss Calib rate b utton Place ce rtifie d trace ab le the rmome te r ...

Страница 2: ... DAMAGES DIRECT OR INCIDENTAL ARISING OUT OF OR RELATED TO THE USE OF THIS MANUAL For repair information or replacement parts assistance from the manufacturer call Technical Services using our tollfreetelephonenumber 800 438 4851 FAX 740 373 4189 NOTE THE 230V UNITS DESCRIBED IN THIS MANUAL WERE DESIGNED SPECIFICALLY FOR THE EUROPEAN MARKET AND ARE SUPPLIED WITH A EUROPEAN STYLE POWER CORD FOR DOM...

Страница 3: ... F NOV01 APR05 JUN06 SEPT08 APR09 Add caution acidic caustic substance newmanuals manufacturelocation consolidated with 3177691 230V removed CE symbol and declaration Revised water specifications Added UL required info 5 36100110 34001605 ECR23443 23496 ECR24808 ECR25448 BA 670 ...

Страница 4: ...UCTION 1 UNPACKING AND DAMAGE 1 GENERAL INFORMATION 2 PERFORMANCE 3 INSTALLATION 3 EXPLANATION OF CONTROLS 4 OPERATION 5 MAINTENANCE 6 TROUBLESHOOTING 7 REPLACEMENT PARTS LIST 10 ASSEMBLY SCHEMATIC DWGS 11 WARRANTY 15 ...

Страница 5: ......

Страница 6: ...packingmaterialuntilunitisput into service This merchandise was carefully packedandthoroughlyinspectedbeforeleaving our factory 2 02 Responsibility for safe delivery was assumedbythecarrieruponacceptanceofthe shipment therefore claimsforlossordamage sustained in transit must be made upon the carrier by the recipient as follows 1 Visible Loss or Damage Note any external evidence of loss or damage o...

Страница 7: ...RAL DEPOSITS AND TO PREVENT CORROSION USE ONLY DISTILLED WATER IN THE BATH WARNING USING CHLORINATED TAP WATER OR ADDITIVES THAT CONTAIN CHLORINE WILL VOID THE MANUFACTURER S WARRANTY SIMILARLY HIGH PURITY DEIONIZED WATER THAT DOES NOT FALL WITH IN RESISTIVITY RANGE OF 50K TO 1M OHM AND PH RANGE OF 7 TO 9 WILL ALSO VOID THE WARRANTY CONTACT TECHNICAL SERVICES WITH ANY QUESTIONS 3 03 The proportion...

Страница 8: ...fts ventilating outlets radiators and other rapidly changingambientconditions Environmental Conditions This instrument is designed to operate safely underthefollowingconditions IndoorUseOnly Temperature 5 to 40 C MaximumRelativeHumidity 80 for temperatures to 22 C Maximum Altitude 2000 meters Maximum performance is assured across the followingtemperaturerange 15ºC to 45ºC 5 02 Electrical Connectio...

Страница 9: ...l used to prevent thermalrunawayintheeventthattheprimary controlfails 6 03 Control Panel The Control Panel provides a digital readout of temperature readings heater ON indicator and pro vides keys for user to make their desired selections 1 LED Display Four digits are used to display the actual set and calibrated valuesfortemperature 2 Heater On Lamp The HEATER ON lamp is illuminated when power is...

Страница 10: ...will read actualbathtemperature 3 ToSelecttheOperatingTemperature Model 2860 has four user selectable temperature setpoints of 35 0 41 5 44 5 and45 5 C Toselectthedesiredoperating temperature presstheVordownWbutton until the desired temperature is displayed PressEnterbuttontoacceptthenewsetpoint Thedisplaywillflash indicatingthatthenew settinghasbeenentered Whentheflashing stops the display return...

Страница 11: ...ome discolored by iron rust use the following procedure to remove all traces of the rust and restore the stainless steel WARNING OBSERVE THE FOLLOWING SAFETY PRECAUTIONS USE HEAVY GLOVES OR OTHER ADEQUATE HAND PROTECTION WEAR GOGGLES OR OTHER ADEQUATE EYEPROTECTION ONLYWORKINAREAS WITH ADEQUATE VENTILATION Prepare a solution of 20 nitric and 1 5 hydrochloric acid if preferred a 2 to 5 solution of ...

Страница 12: ...it lamp is off If light is on turn High Limit control fully clockwise 3 Check temperature probe A Disconnect unit from electrical supply B Disconnect temperature probe connector J302 from CPU board C Apply a 5VDC to the temperature probe 5 to pin 1 of blue connector Ground to pin 3 of blue connector The connector has these numbers embedded on it D Place a voltmeter between pins 2 3 on temperature ...

Страница 13: ... See Paragraph 9 03 Step 6 5 Check TRIAC Q2 on power supply board A Turn power on B Let bath stabilize to desired setpoint value C Place an AC voltmeter between pins 1 5 on wire side of connector J105 on power supply board Voltmeter should fluctuate between 0 volts and anything less than 5 volts D If voltmeter does not show specified values in Step C replace power supply board 9 03 No Heat 4 Check...

Страница 14: ...pump 9 07 Replace CPU or Display Keyboard board 1 The software must be configured to match the Bath model To enter configuration mode press the upper left push button switch while applying power The upper left switch is labeled Calibrate This switch is used to enter configuration codes and steps through the different levels of configurations Immediately following power up with the switch depressed...

Страница 15: ...ssembly Display 3176679 Wire Harness 3177620 3173909 Temp Probe Assembly Plug w Jack 3167038 Probe Thermometer Holder 3177444 Bath O Ring Kit 3167129 Heater Kit 3167160 3175478 Heater Thermostat Bracket N A 3174465 High Limit Thermostat 3175458 Hi Limit Light Amber 3177575 3173907 On Off Switch 3175318 Line Cord 3178034 319481 Diffuser Plate 3163799 Gable Cover 3166219 Power Supply Fuse Mains N A ...

Страница 16: ...11 Stir Pump Assembly Stir Pump Extension On Off Power Switch Diffuser Plate 11 ASSEMBLY AND SCHEMATIC DRAWINGS FRONT SECTION VIEW ...

Страница 17: ...nsulation Certified Traceable Thermometer Holder Thermometer not supplied Temp Sensor Assembly Mounted Installed Here Stir Pump Knob Main Line Power Cord Power Line to Stir Pump Temp Probe Jack Plug probe end into jack Stir Pump Bracket ...

Страница 18: ...able Thermometer Holder Thermometer not supplied Temp Sensor Assembly Mounted Installed Here Stir Pump Knob Main Line Power Cord Receptacle Power Line to Stir Pump Temp Probe Jack Plug probe end into jack Stir Pump Bracket 230V UNIT BACK VIEW ...

Страница 19: ...14 TOP VIEW Temp Sensor Assembly Holder CertifiedTraceable ThermometerHolder Thermometernotsupplied Stir Pump Assembly Stir Pump Bracket Bath Pan Tub ...

Страница 20: ...15 ELECTRICAL DIAGRAM ...

Страница 21: ... warranty period The Technical Services Department must give prior approval for return of any components or equipment At Thermo s option all non conforming parts must be returned to Thermo Electron Corporation postage paid and replacement parts are shipped FOB destination THIS WARRANTY IS EXCLUSIVE AND IN LIEU OF ALL OTHER WARRANTIES WHETHER WRITTEN ORAL OR IMPLIED NO WARRANTIES OF MERCHANTABILITY...

Страница 22: ...m this warranty Replacement or repair of components parts or equipment under this warranty shall not extend the warranty to either the equipment or to the component part beyond the original warranty period The Technical Services Department must give prior approval for return of any components or equipment At Thermo s option all non conforming parts must be returned to Thermo postage paid and repla...

Страница 23: ......

Отзывы: